TISSUE IMPEDANCE MEASUREMENTS USING THE MICROWAVE NETWORK ANALYZER

Bioeffects Seen

Authors not listed

Share:

The authors describe using a microwave network analyzer to measure tissue dielectric properties from 0.1 to 18 GHz with estimated errors around 2%.

Plain English Summary

Summary written for general audiences

The method uses a modified Roberts-von Hippel technique with a microwave network analyzer to measure the reflection coefficient from a tissue sample confined by a teflon plug in a precision coaxial line. Achieving high accuracy requires quarter-wavelength sample lengths, thickness measurement within 0.001 cm, and calibration against known terminations each time a sample is measured. Tests on distilled water, physiological saline, and methanol showed close agreement with known dielectric values, validating the technique.

Why This Matters

The reflection coefficient is most sensitive to tissue dielectric properties when the sample length is near one-quarter wavelength in the material, where it is nearly real and near a minimum. System errors are modeled as a one-port error network whose S matrix is measured using a precision short circuit, a quarter-wave displaced short circuit, and a sliding 50 ohm load at three positions.

Cite This Study
Unknown (n.d.). TISSUE IMPEDANCE MEASUREMENTS USING THE MICROWAVE NETWORK ANALYZER.
Show BibTeX
@article{tissue_impedance_measurements_using_the_microwave_network_analyzer_g5443,
  author = {Unknown},
  title = {TISSUE IMPEDANCE MEASUREMENTS USING THE MICROWAVE NETWORK ANALYZER},
  year = {n.d.},
  
  
}
No DOI on file for this study.

Quick Questions About This Study

The method measures tissue dielectric properties between 0.1 and 18 GHz.
The requirements are quarter-wavelength sample lengths, sample thickness measured to within 0.001 cm, and calibration against known terminations for each tissue measurement.
The technique was tested on distilled water, physiological saline, and methanol, and results agreed closely with known dielectric values for these liquids.
System errors include directivity errors in the directional couplers, residual reflections from the APC-7 connectors, and reflections from the teflon plug. These errors can change each time the teflon plug is moved or connections are changed, so the calibration must be repeated for each tissue measurement.